Accuracy of SenseWear Pro2 Armband to predict resting energy expenditure in childhood obesity

Author(s): Predieri B, Bruzzi P, Lami F, Vellani G, Malavolti M, et al.

Abstract

Objective: The accuracy of the SenseWear Pro2 Armband (SWA) in estimating resting energy expenditure (REE) in children and adolescents with obesity, using indirect calorimetry (IC) as a reference was evaluated.

Design and methods: REE was assessed using both the SWA and IC in 40 obese subjects (26 M/14 F, age 11.5 ± 2.57 years, z-score BMI 3.14 ± 0.53). The agreement between methods was assessed by the Bland-Altman procedure. The relationship between REE assessments and patients' characteristics was also analyzed.

Results: SWA- and IC-derived estimates of REE showed a significant correlation (r = 0.614; P < 0.001), but the SWA overestimated mean REE by 13% (P < 0.001). Age and kilogram of fat-free mass (kg-FFM) were significantly correlated with both REE estimation by SWA (r = 0.434 and r = 0.564, respectively) and IC (r = 0.401 and r = 0.518, respectively). Only kg-FFM was demonstrated to be the main predictor factor of REE variability (r(2) 79% SWA; 75% IC).

Conclusions: The SWA overestimated mean REE in childhood obesity, suggesting that the SWA and IC are not yet interchangeable methods. This would require improving the SWA by developing better algorithms for predicting REE and, probably, bias in each individual REE could be reduced by an adjustment for subjects' kg-FFM.

Similar Articles

Pregnancy weight gain: still controversial

Author(s): Abrams B, Altman SL, Pickett KE

Interventions for preventing excessive weight gain during pregnancy

Author(s): Muktabhant B, Lumbiganon P, Ngamjarus C, Dowswell T.

Gestational weight gain in overweight and obese women enrolled in a healthy lifestyle and eating habits program

Author(s): Petrella E, Malavolti M, Bertarini V, Pignatti L, Neri I, et al.

Objectively recorded physical activity in early pregnancy: A multiethnic population-based study

Author(s): Berntsen S, Richardsen KR, Mørkrid K, Sletner L, Birkeland KI, et al.

[Mediterranean diet: not only food]

Author(s): da Vico L, Agostini S, Brazzo S, Biffi B, Masini ML

Fluid changes during pregnancy: use of bioimpedance spectroscopy

Author(s): Van Loan MD, Kopp LE, King JC, Wong WW, Mayclin PL

Body composition during normal pregnancy: reference ranges

Author(s): Larciprete G, Valensise H, Vasapollo B, Altomare F, Sorge R, et al.

A new device for measuring resting energy expenditure (REE) in healthy subjects

Author(s): Malavolti M, Pietrobelli A, Dugoni M, Poli M, Romagnoli E, et al.

Evaluation of a multisensor armband in estimating energy expenditure in obese individuals

Author(s): Papazoglou D, Augello G, Tagliaferri M, Savia G, Marzullo P, et al.

Compendium of physical activities: an update of activity codes and MET intensities

Author(s): Ainsworth BE, Haskell WL, Whitt MC, Irwin ML, Swartz AM, et al.

Total body water in pregnancy: assessment by using bioelectrical impedance

Author(s): Lukaski HC, Siders WA, Nielsen EJ, Hall CB

Bioelectrical impedance analysis during pregnancy and neonatal birth weight

Author(s): Ghezzi F, Franchi M, Balestreri D, Lischetti B, Mele MC, et al.

Multifrequency bioelectrical impedance analysis in women with a normal and hypertensive pregnancy

Author(s): Valensise H, Andreoli A, Lello S, Magnani F, Romanini C, et al.

Nifedipine-induced changes in body composition in hypertensive patients at term

Author(s): Valensise H, Larciprete G, Vasapollo B, Novelli G, Altomare F, et al.